To keep developing AI medical software, define the changes in advance
An AI model is retrained, and every change can trigger a fresh review by the notified body. The AI Act offers a way out: changes that the provider pre-determines and describes at the initial conformity assessment do not count as a substantial modification. The Medical Device Coordination Group applies this to the MDR, and the FDA knows the principle as a Predetermined Change Control Plan. This article sets out what applies, what remains open and what you can prepare today.
An AI model in medical software is rarely finished when it is certified. New data becomes available, a scanner from another manufacturer delivers different images, performance in one patient group is to be improved. Under the MDR, each of these changes may require approval from the notified body, and under the AI Act a new conformity assessment. Teams that only consider this after certification end up planning their development around the review cycle.
The basis: two regulations, two notions of change
The MDR requires manufacturers to have procedures that take changes in device design or characteristics into account in a timely manner (Article 10(9)). Where the notified body has assessed the technical documentation and issued a certificate, changes require its approval if they could affect safety and performance or the conditions prescribed for use. The manufacturer informs the notified body of the planned change, and the notified body decides whether a supplement to the certificate suffices or a new conformity assessment is needed (Annex IX Section 4.10). Planned substantial changes to the quality management system or to the device range covered must also be notified (Annex IX Section 2.4).
The AI Act brings its own concept: the substantial modification. This is a change after placing on the market that was not foreseen or planned in the initial conformity assessment and that affects compliance with the requirements for high-risk AI systems or modifies the intended purpose (Article 3 point (23)). A high-risk AI system that undergoes a substantial modification is subject to a new conformity assessment (Article 43(4), first sentence).
For AI in medical devices classified as high-risk under Article 6(1), these obligations apply from 2 August 2028. The original date was 2 August 2027; the Digital Omnibus on AI, Regulation (EU) 2026/1744, has postponed it.
The way out: pre-determine the changes
The second sentence of Article 43(4) of the AI Act contains an exception. For high-risk AI systems that continue to learn after being placed on the market, changes to the system and its performance do not constitute a substantial modification if the provider pre-determined them at the moment of the initial conformity assessment and they are described in the technical documentation. There, Annex IV point 2(f) requires a detailed description of the pre-determined changes and of the technical solutions that ensure continuous compliance. Point (g) expressly requires test logs and test reports for these changes as well.
In June 2025, the Medical Device Coordination Group (MDCG) and the EU Artificial Intelligence Board published a joint guidance document, MDCG 2025-6. According to it, pre-determined changes assessed at the initial conformity assessment should also not be understood as a change to the certified medical device under Annex IX Section 4.10 of the MDR (question 30). The rule belongs in the manufacturer's change management procedure and in the technical documentation. The changes must be clearly specified at the conformity assessment, together with the initial performance expectations and the mechanisms for validating and managing changes after placing on the market. The guidance is not legally binding; it was endorsed by two bodies in which all Member States are represented.
For comparison: the FDA's Predetermined Change Control Plan
In the United States, the principle has had a statutory basis since 2022 (Section 515C of the Federal Food, Drug, and Cosmetic Act). The FDA published final guidance for AI-enabled software in December 2024 and updated it in August 2025. An authorised Predetermined Change Control Plan (PCCP) allows a device to be modified in line with the plan, without separate prior authorisation for each significant change. It has three parts:
- Description of Modifications: which changes are planned, such as retraining on new data or new compatible input sources.
- Modification Protocol: how each change is developed, validated and implemented, with pre-defined acceptance criteria.
- Impact Assessment: the benefits and risks of each individual change and of all changes taken together.
The FDA guidance covers models that are modified automatically as well as models that are retrained and released by people. Changes in the plan must keep the device within its intended use.
How to proceed
- Outline the development of the coming years before submission: which changes are foreseeable, and which are deliberately excluded?
- For each planned change, define the data, test method, acceptance criteria and the route back to the previous version. In our experience, the precision of the acceptance criteria decides whether a plan holds up under review.
- Integrate the plan into the change management procedure of the quality management system and keep it in the technical documentation at the place foreseen by Annex IV point 2(f).
- Engage with the notified body early. In our assessment, whether and how it includes the plan in its review has not yet been settled consistently.
- If you also market in the United States, we consider it sensible to align the plan with the structure of the FDA guidance, so that both markets rest on a common basis.
Where this does not apply
The route has limits. Under the AI Act's definition, a change to the intended purpose is always a substantial modification; no plan can anticipate it. Changes not covered by the plan follow the normal route under the MDR and the AI Act.
It is open which systems the exception covers. The wording of Article 43(4) refers to systems that continue to learn after being placed on the market, and MDCG 2025-6 adopts this wording. Whether it also covers a model that the manufacturer retrains and releases at fixed intervals, as the FDA expressly provides for, is in our assessment left open by the text. Under Article 96(1)(c), the Commission is to develop guidelines on substantial modification; MDCG 2025-6 announces that their application to medical devices will then be assessed, and points to the work of the International Medical Device Regulators Forum (IMDRF) as the likely basis for future guidance on change control plans. We were not aware of any EU guidance specifically on change control plans for medical devices at the time of writing.
There is also the Commission's proposal of December 2025 to simplify the MDR. The legislative procedure is ongoing; its effect on change management can only be judged once the text is adopted. Finally, the AI Act exception only concerns high-risk AI systems. Software that does not involve a notified body remains subject to the general obligations of the MDR. This article does not constitute legal advice.
Conclusion
For AI medical software that is meant to keep evolving, the effort required for later changes is largely decided at the first certification. A clearly described change plan with sharp acceptance criteria is not yet a standardised instrument in the EU, but it is the best preparation for what the AI Act will require from August 2028.
Sources: Regulation (EU) 2024/1689 (Artificial Intelligence Act), Article 3 point (23), Article 43(4), Article 96(1)(c) and Annex IV point 2(f) and (g). Regulation (EU) 2026/1744 (Digital Omnibus on AI), OJ L of 24 July 2026, Article 1 point (40) (amending Article 113 of the AI Act). Regulation (EU) 2017/745 (Medical Device Regulation, MDR), Article 10(9) and Annex IX Sections 2.4 and 4.10. Medical Device Coordination Group and Artificial Intelligence Board: MDCG 2025-6 / AIB 2025-1, Interplay between the Medical Devices Regulation & In vitro Diagnostic Medical Devices Regulation and the Artificial Intelligence Act, June 2025 — questions 29 and 30. U.S. Food and Drug Administration: Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions, Guidance for Industry and FDA Staff, originally issued 4 December 2024, updated 18 August 2025 — Sections III, V and VI; Federal Food, Drug, and Cosmetic Act, Section 515C. European Commission: Proposal on simplifying the rules on medical devices and in vitro diagnostic medical devices, COM(2025) 1023, 16 December 2025 — procedure not concluded at the time of writing.